2.
Measurement Process Characterization
2.3. Calibration 2.3.4. Catalog of calibration designs
|
|||
Purpose | The purpose of this section is to explain why calibration of angle blocks of the same size in groups is more efficient than calibration of angle blocks individually. | ||
Calibration schematic for five angle blocks showing the reference as block 1 in the center of the diagram, the check standard as block 2 at the top; and the test blocks as blocks 3, 4, and 5. |
A schematic of a calibration scheme for one reference block, one
check standard, and three test blocks is shown below.
The reference block, R, is shown in the
center of the diagram and the check standard, C,
is shown at the top of the diagram.
![]() |
||
Block sizes |
Angle blocks normally come in sets of
1, 3, 5, 20, 30 minutes 1, 3, 5, 15, 30, 45 degrees and blocks of the same nominal size from 4, 5 or 6 different sets can be calibrated simultaneously using one of the designs shown in this catalog. |
||
Restraint | The solution to the calibration design depends on the known value of a reference block, which is compared with the test blocks. The reference block is designated as block 1 for the purpose of this discussion. | ||
Check standard | It is suggested that block 2 be reserved for a check standard that is maintained in the laboratory for quality control purposes. | ||
Calibration scheme | A calibration scheme developed by Charles Reeve (Reeve) at the National Institute of Standards and Technology for calibrating customer angle blocks is explained on this page. The reader is encouraged to obtain a copy of the publication for details on the calibration setup and quality control checks for angle block calibrations. | ||
Series of measurements for calibrating 4, 5, and 6 angle blocks simultaneously | For all of the designs, the measurements are made in groups of seven starting with the measurements of blocks in the following order: 2-3-2-1-2-4-2. Schematically, the calibration design is completed by counter-clockwise rotation of the test blocks about the reference block, one-at-a-time, with 7 readings for each series reduced to 3 difference measurements. For n angle blocks (including the reference block), this amounts to n - 1 series of 7 readings. The series for 4, 5, and 6 angle blocks are shown below. | ||
Measurements for 4 angle blocks |
Series 1: 2-3-2-1-2-4-2 Series 2: 4-2-4-1-4-3-4 Series 3: 3-4-3-1-3-2-3 |
||
Measurements for 5 angle blocks (see diagram) |
|
||
Measurements for 6 angle blocks |
Series 1: 2-3-2-1-2-4-2 Series 2: 6-2-6-1-6-3-6 Series 3: 5-6-5-1-5-2-5 Series 4: 4-5-4-1-4-6-4 Series 5: 3-4-3-1-3-5-3 |
||
Equations for the measurements in the first series showing error sources |
The equations explaining the seven measurements for the first series
in terms of the errors in the measurement system are:
with B a bias associated with the instrument, d is a linear drift factor, X is the value of the angle block to be determined; and the error terms relate to random errors of measurement. |
||
Calibration procedure depends on difference measurements |
The check block, C, is measured before and after each test
block, and the difference measurements (which are not the same as
the difference measurements for calibrations of mass weights, gage
blocks, etc.) are constructed to take advantage of this situation.
Thus, the 7 readings are reduced to 3 difference measurements for the
first series as follows:
|
||
Design matrix |
As an example, the design matrix for n = 4 angle blocks is
shown below.
1 1 1 1 0 1 -1 0 -1 1 0 0 0 1 0 -1 0 -1 0 1 -1 0 0 1 0 0 -1 1 0 0 1 -1 -1 0 1 0 0 -1 1 0 The design matrix is shown with the solution matrix for identification purposes only because the least-squares solution is weighted (Reeve) to account for the fact that test blocks are measured twice as many times as the reference block. The weight matrix is not shown. |
||
Solutions to the calibration designs measurements |
Solutions to the angle block designs are shown on the following pages.
The solution matrix and factors for the repeatability standard deviation
are to be interpreted as explained in solutions to
calibration designs . As an example, the solution for the
design for n=4 angle blocks
is as follows:
The solution for the reference standard is shown under the first column
of the solution matrix; for the check standard under the second
column; for the first test block under the third column; and for the
second test block under the fourth column. Notice that the estimate for
the reference block is guaranteed to be R*, regardless of the
measurement results, because of the restraint that
is imposed on the design. Specifically,
|
||
Calibrations can be run for top and bottom faces of blocks | The calibration series is run with the blocks all face "up" and is then repeated with the blocks all face "down", and the results averaged. The difference between the two series can be large compared to the repeatability standard deviation, in which case a between-series component of variability must be included in the calculation of the standard deviation of the reported average. | ||
Calculation of standard deviations when the blocks are measured in two orientations |
For n blocks, the differences between the values for the
blocks measured in the top ( denoted by "t") and bottom
(denoted by "b") positions are denoted by:
|
||
Standard deviations when the blocks are measured in only one orientation |
If the blocks are measured in only one orientation, there is no way to
estimate the between-series component of variability
and the standard deviation for the value of each block is computed as
|