OBSERVATIONS 1 1 1 1 1
Y(1) + -
Y(2) + -
Y(3) + -
Y(4) + -
Y(5) - +
Y(6) - +
Y(7) + -
Y(8) + -
Y(9) - +
Y(10) + -
RESTRAINT + +
CHECK STANDARD + -
DEGREES OF FREEDOM = 6
SOLUTION MATRIX
DIVISOR = 10
OBSERVATIONS 1 1 1 1 1
Y(1) 2 -2 0 0 0
Y(2) 0 0 0 2 -2
Y(3) 0 0 2 -2 0
Y(4) -1 1 -3 -1 -1
Y(5) -1 1 1 1 3
Y(6) -1 1 1 3 1
Y(7) 0 0 2 0 -2
Y(8) -1 1 -1 -3 -1
Y(9) 1 -1 1 1 3
Y(10) 1 -1 -3 -1 -1
R* 5 5 5 5 5
R* = average value of the two reference weights
FACTORS FOR REPEATABILITY STANDARD DEVIATIONS
WT K1 1 1 1 1 1
1 0.5477 +
1 0.5477 +
1 0.5477 +
2 0.8944 + +
3 1.2247 + + +
0 0.6325 + -
| |||