2.
Measurement Process Characterization
2.3. Calibration 2.3.4. Catalog of calibration designs 2.3.4.2. Drift-elimination designs for gage blocks
|
|||
Doiron 8-16 Design OBSERVATIONS 1 1 1 1 1 1 1 1 Y(1) + - Y(2) + - Y(3) - + Y(4) - + Y(5) + - Y(6) - + Y(7) - + Y(8) - + Y(9) - + Y(10) - + Y(11) + - Y(12) - + Y(13) - + Y(14) - + Y(15) + - Y(16) + - RESTRAINT + CHECK STANDARD + DEGREES OF FREEDOM = 9 SOLUTION MATRIX DIVISOR = 2852 OBSERVATIONS 1 1 1 1 1 1 1 1 Y(1) 0 -1392 -620 -472 -516 -976 -824 -916 Y(2) 0 60 248 -78 96 878 -112 -526 Y(3) 0 352 124 -315 278 255 864 289 Y(4) 0 516 992 470 1396 706 748 610 Y(5) 0 -356 620 35 286 -979 -96 -349 Y(6) 0 92 0 23 -138 253 -552 667 Y(7) 0 -148 -992 335 -522 -407 -104 -81 Y(8) 0 -416 372 113 190 995 16 177 Y(9) 0 308 -248 170 -648 134 756 342 Y(10) 0 472 620 955 470 585 640 663 Y(11) 0 476 -124 -191 -94 -117 -128 -703 Y(12) 0 -104 -620 -150 404 -286 4 -134 Y(13) 0 472 620 955 470 585 640 663 Y(14) 0 444 124 -292 140 508 312 956 Y(15) 0 104 620 150 -404 286 -4 134 Y(16) 0 568 -124 -168 -232 136 -680 -36 R* 2852 2852 2852 2852 2852 2852 2852 2852 R* = value of reference block FACTORS FOR REPEATABILITY STANDARD DEVIATIONS WT FACTOR 1 1 1 1 1 1 1 1 1 0.0000 + 1 0.6986 + 1 0.7518 + 1 0.5787 + 1 0.6996 + 1 0.8313 + 1 0.7262 + 1 0.7534 + 1 0.6986 +Explanation of notation and interpretation of tables |