2.
Measurement Process Characterization
2.3. Calibration 2.3.4. Catalog of calibration designs 2.3.4.2. Drift-elimination designs for gage blocks
|
|||
Doiron 6-12 Design OBSERVATIONS 1 1 1 1 1 1 Y(1) + - Y(2) - + Y(3) - + Y(4) - + Y(5) - + Y(6) + - Y(7) + - Y(8) + - Y(9) + - Y(10) - + Y(11) + - Y(12) - + RESTRAINT + CHECK STANDARD + DEGREES OF FREEDOM = 7 SOLUTION MATRIX DIVISOR = 360 OBSERVATIONS 1 1 1 1 1 1 Y(1) 0 -136 -96 -76 -72 -76 Y(2) 0 -4 -24 -79 72 11 Y(3) 0 -20 -120 -35 0 55 Y(4) 0 4 24 -11 -72 79 Y(5) 0 -60 0 75 0 -15 Y(6) 0 20 120 -55 0 35 Y(7) 0 -76 -96 -61 -72 -151 Y(8) 0 64 24 4 -72 4 Y(9) 0 40 -120 -20 0 -20 Y(10) 0 72 72 72 144 72 Y(11) 0 60 0 15 0 -75 Y(12) 0 76 96 151 72 61 R* 360 360 360 360 360 360 R* = Value of the reference standard FACTORS FOR REPEATABILITY STANDARD DEVIATIONS NOM FACTOR 1 1 1 1 1 1 1 0.0000 + 1 0.6146 + 1 0.7746 + 1 0.6476 + 1 0.6325 + 1 0.6476 + 1 0.6146 +Explanation of notation and interpretation of tables |