2.
Measurement Process Characterization
2.3. Calibration 2.3.4. Catalog of calibration designs 2.3.4.2. Drift-elimination designs for gage blocks
|
|||
Doiron 5-10 Design
OBSERVATIONS 1 1 1 1 1
Y(1) + -
Y(2) - +
Y(3) + -
Y(4) - +
Y(5) - +
Y(6) + -
Y(7) - +
Y(8) + -
Y(9) - +
Y(10) + -
RESTRAINT +
CHECK STANDARD +
DEGREES OF FREEDOM = 6
SOLUTION MATRIX
DIVISOR = 90
OBSERVATIONS 1 1 1 1 1
Y(1) 0 -50 -10 -10 -30
Y(2) 0 20 4 -14 30
Y(3) 0 -10 -29 -11 -15
Y(4) 0 -20 5 5 15
Y(5) 0 0 -18 18 0
Y(6) 0 -10 -11 -29 -15
Y(7) 0 10 29 11 15
Y(8) 0 -20 14 -4 -30
Y(9) 0 10 11 29 15
Y(10) 0 20 -5 -5 -15
R* 90 90 90 90 90
R* = Value of the reference standard
FACTORS FOR REPEATABILITY STANDARD DEVIATIONS
NOM FACTOR
1 1 1 1 1
1 0.0000 +
1 0.7454 +
1 0.5676 +
1 0.5676 +
1 0.7071 +
1 0.7454 +
Explanation of notation and
interpretation of tables
| |||