Doiron 5-10 Design OBSERVATIONS 1 1 1 1 1 Y(1) + - Y(2) - + Y(3) + - Y(4) - + Y(5) - + Y(6) + - Y(7) - + Y(8) + - Y(9) - + Y(10) + - RESTRAINT + CHECK STANDARD + DEGREES OF FREEDOM = 6 SOLUTION MATRIX DIVISOR = 90 OBSERVATIONS 1 1 1 1 1 Y(1) 0 -50 -10 -10 -30 Y(2) 0 20 4 -14 30 Y(3) 0 -10 -29 -11 -15 Y(4) 0 -20 5 5 15 Y(5) 0 0 -18 18 0 Y(6) 0 -10 -11 -29 -15 Y(7) 0 10 29 11 15 Y(8) 0 -20 14 -4 -30 Y(9) 0 10 11 29 15 Y(10) 0 20 -5 -5 -15 R* 90 90 90 90 90 R* = Value of the reference standard FACTORS FOR REPEATABILITY STANDARD DEVIATIONS NOM FACTOR 1 1 1 1 1 1 0.0000 + 1 0.7454 + 1 0.5676 + 1 0.5676 + 1 0.7071 + 1 0.7454 +Explanation of notation and interpretation of tables |