2.
Measurement Process Characterization
2.3. Calibration 2.3.4. Catalog of calibration designs 2.3.4.2. Drift-elimination designs for gage blocks
|
|||
Doiron 4-12 Design OBSERVATIONS 1 1 1 1 Y(1) + - Y(2) + + Y(3) + - Y(4) - + Y(5) + - Y(6) - + Y(7) + - Y(8) + - Y(9) + - Y(10) - + Y(11) - + Y(12) - + RESTRAINT + CHECK STANDARD + DEGREES OF FREEDOM = 9 SOLUTION MATRIX DIVISOR = 8 OBSERVATIONS 1 1 1 1 Y(1) 0 -2 -1 -1 Y(2) 0 1 1 2 Y(3) 0 0 1 -1 Y(4) 0 2 1 1 Y(5) 0 1 -1 0 Y(6) 0 -1 0 1 Y(7) 0 -1 -2 -1 Y(8) 0 1 0 -1 Y(9) 0 -1 -1 -2 Y(10) 0 -1 1 0 Y(11) 0 1 2 1 Y(12) 0 0 -1 1 R* 6 6 6 4 R* = Value of the reference standard FACTORS FOR REPEATABILITY STANDARD DEVIATIONS NOM FACTOR 1 1 1 1 1 0.0000 + 1 0.5000 + 1 0.5000 + 1 0.5000 + 1 0.5000 +Explanation of notation and interpretation of tables |