2.
Measurement Process Characterization
2.3. Calibration 2.3.4. Catalog of calibration designs 2.3.4.2. Drift-elimination designs for gage blocks
|
|||
Doiron 4-12 Design
OBSERVATIONS 1 1 1 1
Y(1) + -
Y(2) + +
Y(3) + -
Y(4) - +
Y(5) + -
Y(6) - +
Y(7) + -
Y(8) + -
Y(9) + -
Y(10) - +
Y(11) - +
Y(12) - +
RESTRAINT +
CHECK STANDARD +
DEGREES OF FREEDOM = 9
SOLUTION MATRIX
DIVISOR = 8
OBSERVATIONS 1 1 1 1
Y(1) 0 -2 -1 -1
Y(2) 0 1 1 2
Y(3) 0 0 1 -1
Y(4) 0 2 1 1
Y(5) 0 1 -1 0
Y(6) 0 -1 0 1
Y(7) 0 -1 -2 -1
Y(8) 0 1 0 -1
Y(9) 0 -1 -1 -2
Y(10) 0 -1 1 0
Y(11) 0 1 2 1
Y(12) 0 0 -1 1
R* 6 6 6 4
R* = Value of the reference standard
FACTORS FOR REPEATABILITY STANDARD DEVIATIONS
NOM FACTOR
1 1 1 1
1 0.0000 +
1 0.5000 +
1 0.5000 +
1 0.5000 +
1 0.5000 +
Explanation of notation and
interpretation of tables
| |||