2.
Measurement Process Characterization
2.3. Calibration 2.3.4. Catalog of calibration designs 2.3.4.2. Drift-elimination designs for gage blocks
|
|||
Doiron 3-6 design
OBSERVATIONS 1 1 1
Y(1) + -
Y(2) - +
Y(3) + -
Y(4) - +
Y(5) - +
Y(6) + -
RESTRAINT +
CHECK STANDARD +
DEGREES OF FREEDOM = 4
SOLUTION MATRIX
DIVISOR = 6
OBSERVATIONS 1 1 1
Y(1) 0 -2 -1
Y(2) 0 1 2
Y(3) 0 1 -1
Y(4) 0 2 1
Y(5) 0 -1 1
Y(6) 0 -1 -2
R* 6 6 6
R* = Value of the reference standard
FACTORS FOR REPEATABILITY STANDARD DEVIATIONS
NOM FACTOR
1 1 1
1 0.0000 +
1 0.5774 +
1 0.5774 +
1 0.5774 +
Explanation of notation and
interpretation of tables
| |||